Service Pack 3 for Release 3.0 ================================= Solved bugs/ limitations/ extensions: System database: The write protection of system database will remove during software initialization There is an update of the current value of SCSI adress (from computer) in the system database during software initialization Microscope: The display of Axioskop 2 FS control is removed Halogen lamp can displayed for the reflection beampath too (for material applications) Configuration Control: For loading of former configurations (before Release 3.0) the reflection turret (FW1) of microscope is considered now (is none for default) If the revolver of the switching unit of NDD moves, there is an update of display of the appropriate emission filter Beam splitter moves only if the accessory channel is active The display of beampathes is correct, if a multitrack consist of NDD-track and conventional track and both tracks use the transmission channel For NDD: All neccessary switching units move in the correct position for the different VIS- and LSM-states Scan Control: The wavelength of used emission filter will be considered for the calculation of Airy diameter of the pinhole It is possible now to move the focus-handwheel during Z stacks with HRZ 200 The calibration of HRZ 200 works correctly Spline scan with more than 4 Channels works Offset and amplifier gain of detector are adjustable now during spot scan Time series Control REUSE of Mean of Roi series works now Stage Control: The message "Speed 2 too large (98->99)!" in the Error Log is removed (only for motorized xy stage) Macros Spot Bleach (espcially for PASCAL) works ________________________________________________ Service Pack 2 for Release 3.0 ================================= Solved bugs/ limitations/ extensions: Main Menu: The names "dye database" and "spectra database" are exchanged Scan to Disk function check now the TEMP environment variable for possible location of temporary image files and all other fixed disks for temporary storage if there is not enough room in the TEMP directory "Scan to disk" works now as effective as for previous Releases Image database: The display of memory for large image databases is corrct now for more then 2 GByte stored images Laser: If the Mai Tai laser is in suspend mode, the "On" button for activation of the laser goes well to switch the laser on Microscope: The LSM tube lens moves correct in the microscope beam path always The display of lamp power of Hal lamp is right now Configuration Control: Work with monitor diode at the LSM 510 without META goes well Without service pack the following problem occurs: after loading a configuration the Non Descanned Detection option was activ Without service pack the following problem occurs: The FW 1 (reflector turret) moves to position 1 during the changes from Lambda Mode to Channel Mode (only for LSM 510 META) Without service pack the following problem occurs: Sometimes reuse in channel mode with multiple tracks required very long time and moved pinholes multiple times back and forth Scan Control: The display of airy units for pinhole is updated now after change of objectives The Zoom slider is limited at 40 now The Scan Time for bidirectional scanning for Lambda Tracks is correct (equal to unidirectional scanning) Auto Z Correction works after modification of the scan mode Modification of z-Interval for HRZ 200 will be considered for the first scan after modification Stage & Focus Control: Fast Z Line mode for the XZ images for Axioskop 2 FS is correct now The tile scan image works without jumps of the stage now Tiem Series: Time series with a bigger size than the RAM will scanned now Macros: At leave the MCS macro the beam path goes in LSM position now There is a new macro available for PASCAL (called "TimeSeriesShutter.lvb"): Possibility to close the laser shutter during long time series (in the time delays) Objectives: Delete of the empty position is possible Add of "empty position" is possible now in the favourite objective list Adjustment of parfocality works correctly now Auto B&C: Set Find for channel 1 works now The check box "Set "Amplifier Gain to 1" works correct now Images: Content of image window can exported now also for VisArt-displays Import of the newest .zvi-flies is possible now Import of some tiff files generated with Photoshop gave wrong image intensities - is fixed Export of gallery sometimes resulted in image files with wrong dimensions- is fixed Offline software: Change of wavelength for tunable laserline is possible now ConfoCor 2: The current date will shown now in the info sheet Measurement with sample carrier in "Optimize Method" did not write the chamber coordinates to the result list - is fixed New Lock-Button for cross hair postion selection is available now Diagram for pinhole adjustent sometimes displayed a graph outside of the axis boundaries - is fixed Button for large camera display is remove Collimator position is now displayed in pinhole adjustement dialog even if the position has not been modified Display of the current carrier position in the "Optimize Method" dialog will be updated when the user moved the stage with buttons or joystick Laser light was on immediately after the start of a measurement in multiple chambers of a carrier - is fixed Collimator position displayed in "Options-FCS Collimator" dialog is always the actual position