Non-ZEISS Direct Interfaces (NZDI)

ZEISS CALYPSO on other CMMs.

ZEISS CALYPSO NZDI provides a uniform software platform for all of your machines, reduces training requirements, minimizes work when moving ZEISS CALYPSO measurement plans from one system to another, and simplifies measurement reporting.

Contact us

CIMCORE CMMs

  • System Requirements

    Supported CIMCORE systems:, Infinite 1.0, Infinite 2.0, Stinger (4000), 1000, 3000

    Supported CIMCORE Probe Systems: Hard Probe, TP2, TP200

    Condition of the CMM: It should be calibrated and within OEM accuracy specifications, as ZEISS CALYPSO will read existing and future calibration data. It should also be in proper working order. We recommend you use your existing calibration service.

    Hardware Not Supported (not inclusive)

    • Laser Line Scanner
    • Hook Probe
    • Tube Probe
  • CALYPSO Requirements
    CALYPSO Versions Supported:
    • CALYPSO 2015, SP5 minimum
    • CALYPSO 2014, SP1 minimum
    • CALYPSO 5.6, SP2 minimum
    CALYPSO NZDI CIMCORE Software Driver
    CIMCORE OEM Software Driver
     
    CALYPSO Licenses Required
    • CALYPSO Manual
    • nonCZcontrol
    Recommended PC:
    • MS Windows 7 64-bit OS
    • Intel Xeon E5-1603 2.8 10M 1066 4C CPU
    • 16GB DDR3-1600 ECC (4x4GB) RAM
    • NVIDIA Quadro 2000 1GB GFX Graphics
    • 256GB SATA 1st SSD
    • 16X SuperMulti DVDRW SATA 1st ODD
    • Intel CT x1 PCIe NIC
    • 1 RS232 serial port only for CIMCORE RS232 systems
    Operating Systems Supported:
    • MS Windows 7 64-bit
    • MS Windows 7 32-bit
  • Installation Instructions

    Step 1: Install CALYPSO software and any service packs prior to installing the interface.

    Step 2: Download and install the CALYPSO NZDI Driver. Follow instructions prompted during the installation.

    Step 3: Install the OEM CIMCORE device driver supplied by CIMCORE.

    For CIMCORE Infinite, Stinger, 1000 and 3000, minimum version WINRDS 4.0 is required

    Step 4: Move configuration files from OEM source to CALYPSO computer

    1. From the CALYPSO computer open CimCore Arm Utilities.
    2. Select Disconnect under Connect menu.
    3. Select Get Armspecs from Disk Screen.
    4. Select the \armdata folder from the hard drive or from the ArmSpecs USB Flash drive.

     

    Step 5: Configure CALYPSO and connect to the NZDI Device.

    Details are found in CALYPSO configuration for NZDI.pdf

  • Calibration Instructions

    By design, the installation of ZEISS CALYPSO on non-ZEISS OEM machines does not require a machine calibration or acceptance test. Machine correction files for machine accuracy are managed and adjusted by the OEM and reside on the OEM Computer. Copying these files to a CALYPSO installation folder will allow CALYPSO to read and apply this calibration data. For quality assurance reasons, you may choose to perform an acceptance test of ZEISS CALYPSO and the CMM system. However, if machine accuracy or repeatability is not within acceptable limits, adjustments must be made to original calibration file by the OEM and moved to the correct CALYPSO installation folder.

    After the installation of CALYPSO software, or after adjustments are made during a machine calibration, move the configuration files from OEM source to CALYPSO computer.

    1. From the CALYPSO computer open CimCore Arm Utilities. 
    2. Select Disconnect under Connect menu. 
    3. Select Get Armspecs from Disk Screen.
    4. Select the \armdata folder from the hard drive or from the ArmSpecs USB Flash drive.
       
We use cookies on this site. Cookies are small text files that are stored on your computer by websites. Cookies are widely used and help to optimize the pages that you view. By using this site, you agree to their use. more