INTRABEAM 600

ZEISS INTRABEAM Workplace

Total weight with full configuration

max. 155 kg (341.72lbs)

Dimensions

900 mm x 1690 mm x 600 mm / 35.43" x 66.54" x 23.62"
(width x height x depth)

Nominal voltage ranges, switchable

(115 V) 110 V – 125 V
(230 V) 220 V – 240 V

Rated frequency

50 Hz – 60 Hz

Connected load

300 VA

Control Console 600

Weight

4.50 kg (9.92lbs)

Dimensions

305 mm x 89 mm x 381 mm / 12.00" x 3.50" x 15.00"
(width x height x depth)

Power connection

Via workplace isolating transformer

Input voltage

100 V - 240 V

Rated frequency

50 Hz – 60 Hz

Power consumption

60 VA

Selectable operating parameters:

(Accelerating voltage / Beam current)

40 kV / 40 μA
50 kV / 5, 1 0, 20 or 40 μA
is the maximum current of the X-ray tube over the entire voltage range.

NC32 INTRABEAM Floor Stand

Rated voltage

100 V / 115 V / 230 V

Rated frequency

50 - 60 Hz

Power consumption

max. 400 VA

Electrical standard

IEC 60601-1;
CAN/CSA-C22.2 No. 601.1-M90

Product classification

Type B

Case protection

IPX0

Protection class

Protection class I

Notice: The information provided shows only an extract of the instructions for use. For additional information please consult the appropriate instructions for use in full.

Downloads

ZEISS INTRABEAM 600
Technical specifications

pages: 24
file size: 1,500 KB

Make a sales request

Any questions about buying our products?

Click here

to get direct assistance from our service team.

Stay up to date

Sign up for updates on current topics, news and events.