Product Detail

ZEISS QUATERA 700

Technical specifications

Technical data

Aspiration

Type of aspiration pump

QUATTRO Pump®

Aspiration control mode

Vacuum control mode, flow control mode

Vacuum range

0 to 700 mmHg
Note that the maximum available vacuum pressure might be limited by the environmental pressure.

Flow rate range

0 to 120 cc/min

Infusion

Type of infusion

Quattro controlled, forced infusion

IOP range

30 - 120 mmHg

Phacoemulsification

Type of handpiece

Piezoelectrical

Frequency

40 kHz (+4.5 kHz/-0.5 kHz)

U/S operation modes

Continuous, pulse, single burst, multi burst, continuous burst, APM

Diathermy

Type

Bipolar generator - generator stops when no HF power is required

Type of handpiece

Handpieces for bipolar diathermy, e.g. diathermy forceps, diathermy pencil eraser

Control mode

Fixed or "Linear" mode by pressing the foot pedal.

Vitrectomy

Type of handpiece

23G Anterior Vit Cutter, pneumatic cutting device with guillotine function (back-and-forth)

Actuation medium

Compressed air from internal source

Cutting rate range anterior vitrectomy

30 - 5000 cuts/mi n (± 20 %)

Dimensions and weight

Height

1835 mm

Width

556 mm

Depth

758 mm

Tray arm length
(from tray front to console)

875 mm

Weight

160 kg

Electrical data

Rated voltage

100 V - 240 V A C (± 10%)

Rated frequency

50 HZ - 60 HZ

Power consumption

1200 VA