ZEISS VISUMAX 800

Technical Specifications

Technical data

Laser type:
femtosecond laser
Available treatment options:
Flap, SMILE® pro, CIRCLE, ICR, Keratoplasty1
Digital assistant systems:
Centration aid CentraLign®
Cyclotorsion adjustment OcuLign®
Nomogram tool VISULYZE®5
Optical data
Maximum laser repetition frequency:
2 MHz
Wavelength:
1043 nm
Pulse duration:
220–580 fs
Surgical microscope
Magnification:
0.7x
Factors for magnification change:
0.4 / 0.6 / 1.0 / 1.6 / 2.5
Eyepiece magnification:
10x
Filters:
blue, barrier filter (yellow)

Slit illumination:

slit width ≤ 0.3 mm / 0.7 mm
slit length 11.0 mm
Ambient conditions for operation
Temperature:
+18 °C to +25 °C
Humidity:
30 % to 70 %
Measurements
Device mass:
520 kg
Minimum floor load:
2.5 kN/m1
Footprint standalone:
L x W:1710 mm x 925 mm

Downloads

    • Refractive Workplace OnePager EN

      Pages: 2
      File size: 308 KB
    • VISULYZE Onepager EN

      Pages: 2
      File size: 198 KB
    • VISUMAX 800 Brochure EN

      Pages: 7
      File size: 2 MB
    • VISUMAX 800 Case of the Month Dr. Wiltfang EN

      Pages: 1
      File size: 408 KB
    • VISUMAX 800 Case of the month Prof. Sekundo EN

      Pages: 1
      File size: 1 MB
  • 1

    Data on file, verification report

  • 2

    Data on file, myopia with optical zone 6.5 mm

  • 3

    Data on file, flap diameter ≤ 8.0 mm, spot distance 4.5 µm, track distance 2 µm

  • 4