ROMER CMMs

  • System Requirements
    Supported ROMER systems: Absolute
     
    Supported ROMER Probe Systems: Hard Probe, TP2, TP200
     
    Condition of the CMM: It should be calibrated and within OEM accuracy specifications, as ZEISS CALYPSO will read existing and future calibration data. It should also be in proper working order. We recommend you use your existing calibration service.
     

    Hardware Not Supported (not inclusive)

    • ROMER Laser Line Scanner
    • Hook Probe
    • Tube Probe
     

  • CALYPSO Requirements

    CALYPSO Versions Supported:

    • CALYPSO 2018 SP3 (6.6.16)
    • CALYPSO 2017 SP2 minimum
    • CALYPSO 2016 SP3 minimum
    • CALYPSO 2015, SP5 minimum
    • CALYPSO 2014, SP1 minimum
    • CALYPSO 5.6, SP2 minimum

    CALYPSO NZDI ROMER Software Driver

    ROMER OEM Software Driver

    CALYPSO Licenses Required

    • CALYPSO Manual
    • nonCZcontrol

    Recommended PC:

    • Windows 10 64-bit Operating System
    • Intel Xeon W-2123 3.6 4C CPU
    • 64GB RAM
    • 256GB SATA SSD (minimum)
    • NVIDIA Quadro P2000 5GB 1st GFX
    • 2 Network Interface Cards
    • 1 RS232 Serial Card

    Operating Systems Supported:

    • CALYPSO 2018 to 2016; MS Windows 10
    • CALYPSO 5.6 to 2018; MS Windows 7 64 bit
  • Installation Instructions

    Step 1: Install CALYPSO software and any service packs prior to installing the interface.

    Step 2: Download and install the CALYPSO NZDI Driver. Follow instructions prompted during the installation.

    Step 3: For ROMER arms, install the OEM ROMER device driver supplied by ROMER. 

    • For ROMER Absolute, minimum version RDS 4.1 is required

    Step 4: Configure CALYPSO and connect to the NZDI Device.

  • Calibration Instructions

    By design, the installation of ZEISS CALYPSO on non-ZEISS OEM machines does not require a machine calibration or acceptance test. Machine correction files for machine accuracy are managed by the OEM and reside on the controller. Coordinate data that is sent to ZEISS CALYPSO is considered “calibrated” and is not adjusted or altered. For quality assurance reasons, you may choose to perform an acceptance test of ZEISS CALYPSO and the CMM system. However, if machine accuracy or repeatability is not within acceptable limits, adjustments must be made to the machine and controller using the OEM software as needed. ZEISS Industrial Metrology does not provide this service or machine repair.