MMS 1 |
MMS UV-VIS |
MMS UV |
Optical entrance |
Fiber bundle consisting of approx. 30 Quartz glass fibers with 70 µm core diameter each, designed as cross section converter |
Fiber bundle consisting of approx. 30 Quartz glass fibers with 70 µm core diameter each, designed as cross section converter |
Fiber bundle consisting of approx. 15 Quartz glass fibers with 70 µm core diameter each, designed as cross section converter |
Input round |
Diameter: 0.5 mm NA = 0.22 (has to be filled for full specification) mounted in SMA-coupling |
Diameter: 0.5 mm NA = 0.22 (has to be filled for full specification) mounted in SMA-coupling |
Diameter: 0.4 mm NA = 0.22 (has to be filled for full specification) mounted in SMA-coupling |
Output linear |
70 µm x 2,500 µm (optical entrance) |
70 µm x 2,500 µm (optical entrance) |
70 µm x 1,250 µm (optical entrance) |
Grating |
Flat-field, 366 l/mm (center) |
Flat-field, 366 l/mm (center), blazed for approx. 220 nm |
Flat-field, 1084 l/mm (center), blazed for approx. 220 nm |
Diode array |
Hamamatsu S3904-256Q in a special housing (S5713) (S8381-256Q for MMS 1 NIR enh.), 256 pixels |
Hamamatsu S3904-256Q in a special housing (S5713), 256 pixels |
Hamamatsu S3904-256N in a special housing (S5713), 256 pixels |
Spectral range |
310-1,100 nm specifications for the range 360-900 nm (UV-VIS enhanced) 400-1,000 nm (NIR enhanced) |
190-720 nm specifications for the range 220-720 nm 250-780 nm specifications for the range 250-780 nm |
195-390 nm specifications for the range 220-390 nm |
Wavelength accuracy |
0.3 nm |
0.5 nm |
0.2 nm |
Temperature - induced drift |
< 0.01 nm/K |
< 0.006 nm/K |
< 0.005 nm/K |
Mean spectral pixel pitch |
Δ λ pixel ≈ 3.3 nm |
Δ λ pixel ≈ 2.2 nm |
Δ λ pixel ≈ 0.8 nm |
Resolution |
Δ λ FWHM ≈ 10 nm |
Δ λ FWHM ≈ 7 nm |
Δ λ FWHM ≈ 3 nm |
Sensitivity |
≈ 103 Vs / J |
≈ 103 Vs / J |
≈ 103 Vs / J |
Straylight |
≤ 0.8 % with Halogen lamp for UV-VIS enhanced as transmission at 450 nm with filter GG 495 ≤ 0.2 % with Halogen lamp for NIR enhanced as transmission at 650 nm with filter RG 695 |
≤ 0.3 % with Deuterium lamp as transmission at 310 nm with NaNO2-solution (50 g/l) |
≤ 0.3 % Deuterium lamp as transmission at 240 nm with NaI-solution (10 g/l) |
Dimensions |
70 x 50 x 40 mm3 |
67 x 60 x 40 mm3 |
70 x 60 x 40 mm3 |
Cross section converter Minimal bending radius |
240 mm standard, up to 1 m available 40 mm |
240 mm standard, up to 1 m available 40 mm |
240 mm standard, up to 1 m available 40 mm |