    Specific spectroscopic advantages are gained through the use of Zeiss lamp modules; an integrated signal processor controls and guarantees high stability and dynamic response. The lamps are designed with overload and short circuit protection as well as a soft-start feature to increase their service life. Communication of parameters and status with application software eliminates set up concerns and potential confusion due to multiple light sources. Additionally, adjustable voltage and power intensity, to set energy requirements or color temperature, is now available.

    BLX 606 and CLH 606 light sources allow the simultaneous and homogeneous illumination of up to 6 channels.

  • UV-VIS range sources
    UV-VIS range sources

    Deuterium Lamp Cassette:

    • CLD 600

    Xenon Flash Lamp Cassettes:

    • BLX 600 CZ
    • BLX 600 SMA
    • BLX 606
    Type Continous Flash Flash Flash
    Technical Specifications CLD 600 BLX 600 /CZ6   BLX 600 /SMA   BLX 606
    Spectral Range 210 - 640 nm 240 - 800 nm 220 - 800 nm 220 - 800 nm
    Power Consumption 25 W 15 W 15 W 15 W
    Radiation Output 2.5 mW
    0.6 J/ flash 0.6 J/ flash
    0.6 J/ flash
    Lamp Drift 0.30 % RMS          -
            -
             -
    Noise 0.15 % RMS 1,7 % RMS * 1,7 % RMS *
    1,7 % RMS *
    Shutter integrated Shutter n.a. n.a. n.a.
    Max. Shutter Frequency 1Hz n.a. n.a. n.a.
    Fiber Connector 1x SMA 1x ZEISS (6 mm) 1x SMA 6x SMA
    life time 1000 h 5*108 flashes 5*108 flashes 5*108 flashes
             
    * general recommendation for usage of flash lamps            W i t h   d o u b l e   b e a m   c o n f i g u r a t i o n   o n l y!  
  • VIS-NIR range sources

    VIS-NIR range sources

    Halogen Lamp Cassettes

    • CLH 600
    • CLH 600 F
    • CLH 606 A
    Technical Specifications * CLH 600 CLH 600 F
    CLH 606 A
    Spectral Range 380 - 2400 nm 320 - 2400 nm 380 - 2400 nm
    Lamp Drift 0.3 % RMS 0.3 % RMS 0.3 % RMS
    Noise 0.09 % RMS 0.10 % RMS 0.35 % RMS
    Shutter integrated Shutter integrated Shutter -
    Fiber Connector 1x SMA 1x SMA 6x SMA
           
    *general remark Specification depends also on the used spectrometer configuration  