Shaping oncology workflows of tomorrow

ZEISS INTRABEAM 700

ZEISS INTRABEAM 700 at a glance

ZEISS INTRABEAM 700 device overview

Components
INTRABEAM SMART
Spherical Applicator

1.5 cm – 5.0 cm diameters available
Individual sterile packaging (single use)

New Radiance™ & New INTRABEAM GUI
New Connectivity Features

Automated applicator recognition
Cloud connectivity
PACS connection

INTRABEAM SMART Stand

Fast & Slow Movement Mode
AutoDrape®
AutoBalance

INTRABEAM 700 (Workplace)

X-ray source (XRS 4)

The electrons emitted by the cathode are accelerated through the drift tube with a maximum difference of a potential of 50 kV onto a gold target. The braking energy of the electrons creates low-energy photons, which are distributed isotropically.

Online dose monitoring

The internal radiation monitor (IRM) detects the part of the low-energy photons emitted in the direction of the cathode and records the dose output in real-time.1 The IRM result is displayed in the GUI of the ZEISS INTRABEAM software so that the user knows which dose is being delivered at any time throughout the treatment.

  • Point-source type X-ray emission
  • Spherical dose distribution around the
    isocenter of the X-ray source (XRS 4)
  • Steep dose gradient (approx. 1/r3) in water
    (soft tissue equivalent)
  • Positional accuracy of delivered dose +/- 1 mm at 40 mm treatment diameter (from isocenter)

ZEISS Tumor Workflow

Reconsider the standard of care.

Downloads



Request a demo!

See what the ZEISS INTRABEAM 700 can do for your operating room

Form is loading...

Personal contact information

If you want to have more information on data processing at ZEISS please refer to our data privacy notice


  • 1

    Subject to appropriate calibration.