Specifications
|
Electrical Data
|
|
|---|---|
|
Rated Voltage at 115V
|
100 V - 240 V
|
|
Rated Voltage at 230V
|
220 V - 240 V
|
|
Power Consumption at 115V
|
300 VA
|
|
Power Consumption at 230V
|
300 VA
|
|
Rated Frequency
|
50 Hz - 60 Hz
|
|
Electrical Standard
|
Complying with IEC 60601-1:2005+A1:2012 andIEC 60601-1-2:2014
Protection class I, degree of protection IP X0 (system cart), IP x6 (foot control panel) |
|
Laser data
|
|
|
Laser class |
3R as per IEC60825-1:2014 and IEC 60825:2007 |
|
Laser power |
1 mW |
|
Wavelength
|
488 nm
|
|
Laser safety range
|
32 nm or more away from the tip of the scanner probe, time base 0.25 seconds
|
|
Recording parameters
|
|
|
Field of view
|
Horizontal: approx. 475 μm
Vertical: approx: 267 μm |
|
Image resolution and frame rate
|
1920 x 1080 pixels (full hD) / 0.75 frames per second
1920 x 270 pixels / 2.35 frames per second |
|
Emission Filters
|
Green band-pass filter (517.5 - 572.5 nm(545/55))
Grenn long-pass filter (>515 nm) Red long-pass filter (>572 nm) Neutral density filter (OD3, i.e. 0.1% transmission) |
|
Dimensions and weights of system cart and monitor
|
|
|
Dimensions (W x H xD)
|
750 x 1685 x 725 mm
|
|
Weight
|
165 kg
|
|
Weight of system incl. transport container
|
approx. 335 kg
|
|
Dimensions and weights of scanner probe
|
|
|
Weight of scanner probe
|
1250 g
|
|
Length of scanner probe shaft
|
150 mm
|
|
Diameter of shaft with sterile sheath
|
5 mm
|
|
Length of cable
|
3.8 m
|
|
Sterile concept
|
|
|
Sterile Drape
|
ZEISS Sterile Sheath for CONVIVO
|